After a transvaginal ultrasound in which a probe cover was used and remained intact, what level of reprocessing is required for the endocavitary transducer before the next patient?
Show answer & explanation
Answer: C
Low-level disinfection with a germicidal wipe, because the intact cover prevented contact: Incorrect. Probe covers can have undetected perforations, so an endocavitary transducer is still treated as semicritical regardless of an intact-appearing cover.
Cleaning with soap and water only, since the transducer touched intact skin: Incorrect. An endocavitary probe contacts mucous membranes, not intact skin; soap and water alone does not meet the standard for semicritical devices.
High-level disinfection after cleaning, because the probe contacts mucous membranes: Correct. Under the Spaulding classification, devices contacting mucous membranes are semicritical and require cleaning followed by high-level disinfection after every use, even with a cover.
Steam sterilization in an autoclave, because the probe enters a body cavity: Incorrect. Ultrasound transducers would be damaged by steam autoclaving, and sterilization is required for critical devices entering sterile tissue, not for transvaginal probes.
Spaulding: critical (sterile tissue) → sterilize; semicritical (mucous membranes, e.g., TV/TR probes) → high-level disinfection; noncritical (intact skin) → low-level disinfection.