While setting up an ECG cart, the technician notices the power cord insulation is cracked near the plug and the ground pin is bent. The patient is waiting. What is the correct action?
Show answer & explanation
Answer: A
Remove the cart from service, tag it, and use another machine: Correct. Damaged insulation or a compromised ground creates a leakage-current and shock hazard; the unit must be tagged out and reported to biomedical engineering.
Straighten the ground pin and record the ECG on battery power: Field repairs are not allowed and battery use does not fix a damaged cord; the device still needs inspection.
Use a two-prong adapter so the plug fits the wall outlet: Defeating the ground removes protection against leakage current and increases shock risk.
Wrap the cracked area with tape and report it after the test: Tape is not an approved repair, and using damaged equipment exposes the patient to electrical hazard.
Damaged cords, missing ground pins or adapters are electrical safety hazards: remove from service, tag, report to biomed and use a safe unit.