During the daily calibration check of a spirometer, a technologist injects a 3.00-L calibration syringe. Under the 2019 ATS/ERS spirometry standard, which measured volume fails the check?
Show answer & explanation
Answer: C
2.94 L: 2.94 L is a 2% error, inside the ±3% tolerance (2.91–3.09 L).
3.05 L: 3.05 L is about a 1.7% error and passes.
3.12 L: Correct. 3.12 L is a 4% error, outside the ±3% limit; the spirometer must be recalibrated or serviced before testing.
3.08 L: 3.08 L is about a 2.7% error, still within ±3%.
2019 ATS/ERS: a 3-L syringe check must read within ±3% (2.91–3.09 L); outside that range, do not test patients until the device is corrected.