Question 1 of 10
Water Treatment
The facility applies ANSI/AAMI/ISO 13959:2014 limits for dialysis water. Based on the report shown, which conclusion is correct?
Monthly water microbiology report| Sample site | Bacteria (CFU/mL) | Endotoxin (EU/mL) |
|---|
| RO product | 3 | <0.03 |
| Loop first outlet | 18 | 0.06 |
| Loop return | 42 | 0.14 |